Working Voltage | 220 VAC switchable, 50/60 Hz, single phase Max power: 1.5 kW |
Glove Box Chamber | Material: Stainless steel 304 with 3 mm thickness Chamebr Dimension: 1200*760*900(mm) |
Chamber Condition | Water concentration: <1 ppm (20 °C, 1 atm) Oxygen concentration: <1 ppm (20 °C, 1 atm) |
Leak Rate | <0.05 vol% / hr |
Airlock Chamber | Antechamber dimension: 385 mm Dia. x 600 mm L Small chamber dimension (for tools): 150 mm Dia. x 300 mm L |
Working Gas | Operating gas: Inert gas such as N2, Ar, He Regenerating gas: For regenerating both oxygen and H2O remover, use a mixture of H2 (5-10%) and operating gas |
Gas Purification System | Automated regeneration process Automated Water and oxygen removal Pipeline: 304 stainless steel, KF40, and KF25 ports |
Ports & Fittings | Total three KF40 ports on the back of the glove box |
Pressure Control System | Automatic pressure control by PLC for the Glove Box Chamber Pressure range: +/- 12 mbar Manual pressure control is allowed via a foot pedal |
Purging System | Automatic purging system |
Control System | PLC control system with 7" touchscreen display Water concentration: 0 - 1000 ppm with an accuracy of 0.1 ppm Oxygen concentration: 0 - 1000 ppm with an accuracy of 0.1 ppm Pressure: -12 to 12 mbar with an accuracy of 0.01 mbar |
Pump | 240 L/m heavy duty double stage rotary vane pump for faster gas exhaust Vacuum rate: 240 L/m |
Gloves | A pair of 9 3/4" butyl glove |